IOL Master

  IOLMaster
Online Instruction Manual V.5
printable version
 
IOL Power Calculations IOL Power Calculations
Zeiss IOLMaster Zeiss IOLMaster
IOLMaster User Manual V.5 IOLMaster User Manual V.5
Keratometer
Calibrate Keratometer Aid
Tips for Keratometer
IOLMaster Lens Constants IOL Master Lens Constants
IOL Master Manual Download IOLMaster Checklist (pdf)
IOL Master Manual Download IOLMaster Manual V.5 (pdf)
Contact Us
Disclaimer
Physician Downloads SIA Calculator
Links
References
Special Thanks

Zeiss IOLMaster Customer Service: 1-877-486-7473 ext. 4030

IOLMaster — Calibrating Keratometer Aid.

Calibration instructions for the keratometer adjustment aid (traffic light display)

  1. If the Advanced Technology Software version 5 has been installed on your device and you wish to use the keratometer adjustment aid (traffic light display), it must first of all be calibrated.
  2. Switch the IOLMaster on.
  3. Wait until all internal system tests have been completed and the Patient Manager appears upon confirming the calibration prompt with OK.
  4. In the Options pull-down menu, click on Setup and use the Program Settings to display the Program dialog. Activate the Adjustment aid here (and Automatic, if you wish).
  5. Create a new patient with New.
  6. Change to keratometer mode by clicking the appropriate icon or pressing the <K> key.
  7. Follow the instructions on the screen.

    The test eye must be clean. If the test eye for the IOLMaster is not available, abort the calibration process with Cancel. The Adjustment aid will then be deactivated in the Program Settings dialog and can no longer be used.

    If the software is unable to perform the calibration (message described under item 10 is repeatedly displayed), the cause may be inadequate room lighting. It is recommended that the front panel and test eye be shielded from direct or lateral light. The best results will be obtained when the examination room is slightly darkened.

  8. Then you will be prompted to repeat the process once again.
  9. If the calibration process was successful, a message to this effect will be displayed. The keratometer adjustment aid (traffic light display) can now be used to position the keratometer at the optimum measurement setting.
  10. If a message appears stating that variance between the calibration values of the two calibrations was too wide, the entire calibration process must be repeated from Step 7 onwards.

 

Cataract Surgery Arizona

IOLMaster Help
TOP

Zeiss IOL Master

Intraocular Lens Power Calculations