doctor-hill.com
Carl Zeiss Meditec IOL Master
Cataract Surgery Arizona

IOLMaster — Axial Length Measurement.

IOLMaster axial length measurement

Activate Axial Length Measurement (ALM) Measurement Mode:

  • Click on the ALM icon, or
  • Press key A, or
  • Press the button on the joystick in Overview (OVW) mode.
IOLMaster axial length measurement   Switching to ALM mode will automatically change the magnification ratio: a smaller section of the eye becomes visible with the reflection of the alignment light and a vertical line (1, Fig. 27).
  • The patient should look at the red fixation point in the center. A cross-hair (3, Fig. 27) with a circle in the middle will appear on the display.
  • Fine-align the IOLMaster so that the reflection of the alignment light (2, Fig. 27) appears within the circle.


IOL Master warning

WARNING: Ask the patient if he or she sees the fixation point. If the patient fails to fixate properly, the visual axis will not be correctly recognized, which may result in measuring errors.

In the case of poor visual acuity/high ametropia (> 4 D) it is advisable to measure through the spectacles. If the procedure is followed correctly, no measuring errors will be produced. Measurements should not be taken while a patient is wearing contact lenses, as it will result in measuring errors.

  • Take the measurement by pressing the button on the joystick.

The corresponding display field next to the video image will show the measured axial length. The video image will be overlaid with a red graph similar to those in ultrasonic measuring instruments. Simultaneously, the axial length and signal-to-noise ratio (SNR) of the measuring signal will be displayed. This value is a gauge of the quality of measurement. Measurements with an SNR between 1.6 and 1.9 appear with an exclamation mark (!) after the measured value and the message "Borderline value!" will appear. For evaluation of the SNR see Post-run editing of axial length measurements.

Readings that deviate from the internally calculated mean value by more than 50 μm are shown in red and marked "multiple peaks".

IOLMaster Arizona   Note
Borderline value! does not necessarily mean that the reading is incorrect and must be rejected. It rather means that all axial length measurements for the eye should be checked for plausibility and consistency, e.g. according to the usual ultrasonic biometry criteria. If the "uncertain" values are determined to concur with the other readings, the readings marked Borderline value! should also be accepted as valid axial lengths.


IOLMaster Arizona Important:
The number of measurements is crucial here. The IOLMaster requires FIVE measurements to be taken. Until then, the message Measure again will thus appear. Only after five measurements are recorded will a mean value be passed on to the IOL calculation and an evaluation enabled.

To obtain consistent results we recommend checking the individual axial length measurements and carrying out further measurements if necessary.

With stronger lens opacities, it may be advisable to defocus the IOLMaster. You may choose a reflection (2, Fig. 27 above) as large as the circle on the display. If measurements are even now impossible, the IOLMaster can be refocused and the reflection shifted to the bottom and/or top margin of the circle on the display by varying the vertical adjustment (turning joystick).

IOLMaster Arizona   Note
Defocusing and shifting the reflection within the circle will have no effect on the result, because interferometric axial length measurement is completely independent of distance.
  • For the next measurement of this eye, press the button in the joystick.
IOL Master warning

WARNING: Up to 20 such measurements per eye may be taken on a single day.

Avoid measurements of eyes with retinal detachment. In such cases, measuring errors cannot be precluded.

As a rule, the axial length should be viewed together with the values for corneal refraction and overall refraction, and checked for plausibility. It is likewise helpful to compare the right and left eyes.

The composite signal is calculated after the fifth measurement. Initially, the individual signals are displayed in red. After a delay of about 1 second the composite signal is then displayed in blue. In addition, the axial length reading determined from this composite signal will appear. The composite signal will be re-calculated after each further individual measurement, and an axial length calculated therefrom. Should a reading deviate from another by more than 0.05 mm, it will be displayed in red and the message Multiple peaks will appear. This indicates that the individual measurements should be scrutinized and the composite signal may need to be post-run edited, as explained in: Post-run Editing of Axial Length Measurements.

Until an axial length can be determined from the composite signal, the word Evaluation! will be displayed below the horizontal line in the list of measurements. If necessary, the potential individual measurement errors must be deleted or the composite signal post-run edited, as the readings obtained will not otherwise be accepted for the IOL calculation and database for optimisation of the lens constants. Until the fourth individual measurement has been taken, the last reading will be highlighted in blue. From the fifth individual measurement onwards, the composite signal is highlighted in blue. The blue highlighting can be moved through the table of individual readings with the aid of the cursor buttons ↑ ↓. In this way the signal curves of the individual measurements can be displayed. Deleted individual measurements can be restored with the shortcut <CTRL> + <Z>. The composite signal can be displayed by clicking on the composite reading.

Error in the display field denotes readings with an SNR smaller than 1.6.

The following plausibility tests are performed with the axial length measurement (AL) from the composite signal:

  • AL < 22mm (indication of short ocular axis, use corresponding biometric formula)
  • AL > 25mm (indication of long ocular axis, use corresponding biometric formula)
When both eyes have been measured, the difference in axial lengths between right and left is also checked. If the latter exceeds 0.3 mm, a message appears to check the readings once again.

If this warning appears, be sure to verify that no pathological changes have occurred in the eye. If necessary, the measurements must be repeated (provided the maximum of 20 measurements per eye and day has not already been reached). Only confirm the warning with OK if you are certain that the readings are plausible. Otherwise, try to determine what has caused the implausible readings. A reference to the displayed plausibility test message will be transferred to the comments box.

The number of measurements of the respective eye taken on this particular day is displayed in the Mode field of the status bar next to "ALM". If the count reaches 20 no further measurements of this eye can be taken on this day. The counter cannot be reset. Deleted readings (see above) do not affect the measurement counter!

IOLMaster axial length measurement
Fig. 28 Video image after axial length measurement

ALM Measurement of Non-Aphakic Eyes

To measure non-aphakic eyes, select the corresponding mode from the AL Settings menu. This special AL mode is displayed in the video image field and will be active until you reset it via the menu. The IOL Master will also be reset to Phakic mode if you change to the patient's other eye or a new patient.

If the axis length of eyes with phakic implants is to be measured, and are not listed in the additional AL functions, the following compensation values according to Dr Wolfgang Haigis of Wurzburg University Clinic, Germany, should be used.

IOLMaster axial length measurement

IOLMaster axial length measurement
Fig. 29 AL settings
 

Every implant, e.g. a phakic IOL, influences the measurement of axial length in PCI biometry. If a phakic lens implant is measured in a normal phakic mode, the result will be slightly elevated. The reading must be corrected, depending on the material used and the center. thickness.

Sample calculation for a phakic implant (Acrysof) with a center thickness of 0.2 mm:

Measured value: 23.51 mm Compensation value: -0.03 mm Correct axial length: 23.51 + (-0.03) = 23.48 mm


IOL Master warning WARNING: Two peaks may appear when measuring pseudophakic eyes and with certain intraocular lenses. The first peak is a side maximum of the IOL, while the second peak is produced by the retina. In this case, manual correction is necessary (see Measuring errors with pseudophakic eyes). It is expedient to measure at a number of different points.

 

IOL Master warning WARNING: Use the psph (pseudophakic) button to calculate secondary piggy-back IOLs. For this purpose, the ACD should be measured by a method other than the IOLMaster and the readings thus obtained entered into the appropriate boxes.