IOL Master

  IOLMaster
Online Instruction Manual V.4
printable version
 
IOL Power Calculations IOL Power Calculations
Zeiss IOLMaster Zeiss IOLMaster
IOLMaster User Manual V.4 IOLMaster User Manual V.4
Axial Length (ALM)
Best Technique
Summary of Results
Interpretation of Signals
Signal Curves
Pseudophakic Eyes
Clear Media
Mild Cataract
Moderate Cataract
Dense Cataract
Zooming Display
Post-run Editing
IOLMaster Lens Constants IOL Master Lens Constants
IOL Master Manual Download IOLMaster Manual V.4 (pdf)
Contact Us
Disclaimer
Physician Downloads SIA Calculator
Links
References
Special Thanks

Zeiss IOLMaster Customer Service: 1-877-486-7473 ext. 4030

IOLMaster — Axial Length Measurement
Pseudophakic Eyes.

Pseudophakic Eyes and False Images

IOL Master Pseudophakic false images

Measuring Errors with Pseudophakic Eyes

Two peaks may appear when measuring pseudophakic eyes and with certain intraocular lenses (e.g. Acysof). The first peak (false) is from the IOL, while the second peak is produced by the retina. In this case, manual correction of the axis length is necessary. It is expedient to measure at a number of different points.

Because the IOL Master is an optical device, reflections off the surface of PMMA, acrylic intraocular, and occasionally silicone lenses, can sometimes produce false images. To avoid this, sample several areas around the reticule using the correct Measurement Technique.

If there is a visible reflection off the surface of an intraocular lens implant, excluding the responsible area will usually give correct results. When the axial length is much shorter than expected, a false image is probably in play. Then, a comparison with the fellow eye typically renders a significant difference, usually greater than 3.00 mm.

 

 

 

 

 

 

Cataract Surgery Arizona

IOLMaster Help
TOP

Zeiss IOL Master

Intraocular Lens Power Calculations