Axial Length (ALM)
Best Technique
Summary of Results
Interpretation of Signals
Signal Curves
Pseudophakic Eyes
Clear Media
Mild Cataract
Moderate Cataract
Dense Cataract
Zooming Display
Post-run Editing
IOL Master Manual Download IOLMaster Checklist (pdf)
IOL Master Manual Download IOLMaster Manual V.4 (pdf)
IOL Master Manual Download IOLMaster Manual V.3 (pdf)
SIA Calculator
Disclaimer
Physician Downloads SIA Calculator
Links
References
Special Thanks

Zeiss IOLMaster Customer Service: 1-877-486-7473 ext. 4030

IOLMaster — Axial Length Measurement
Pseudophakic Eyes.

Pseudophakic Eyes and False Images

IOL Master Pseudophakic false images

Measuring Errors with Pseudophakic Eyes

Two peaks may appear when measuring pseudophakic eyes and with certain intraocular lenses (e.g. Acysof). The first peak (false) is from the IOL, while the second peak is produced by the retina. In this case, manual correction of the axis length is necessary. It is expedient to measure at a number of different points.

Because the IOL Master is an optical device, reflections off the surface of PMMA, acrylic intraocular, and occasionally silicone lenses, can sometimes produce false images. To avoid this, sample several areas around the reticule using the correct Measurement Technique.

If there is a visible reflection off the surface of an intraocular lens implant, excluding the responsible area will usually give correct results. When the axial length is much shorter than expected, a false image is probably in play. Then, a comparison with the fellow eye typically renders a significant difference, usually greater than 3.00 mm.

 

 

 

 

 

 


IOLMaster Help
TOP

Zeiss IOLMaster® Tutorial Version 4

Zeiss IOLMaster®

Optical Biometry

Intraocular Lens Power Calculations


East Valley Ophthalmology
5620 East Broadway Road
Mesa, Arizona 85206

Tel: +1-480-981-6111
FAX: +1-480-985-2426

Arizona's Top Eye Doctors - East Valley Ophthalmology provides this online information for educational and communication purposes only and it should not be construed as personal medical advice. Information published on this website is not intended to replace, supplant, or augment a consultation with an eye care professional regarding the viewer/user's own medical care. East Valley Ophthalmology's disclaims any and all liability for injury or other damages that could result from use of the information obtained from this site. Please read our full Terms, Privacy, Infringement

Eye Doctors for Mesa, Gilbert, Chandler, AZ and surrounding areas.

A few of our featured topics:


TOP

Copyright East Valley Ophthalmology. All rights reserved.